Item shown is representative of range - Catalogue No.:GB-C1706

Item shown is representative of range - Catalogue No.:GB-C1706

Hettich - Rotina 380R

Stock Code: GB-C1706

Centrifuge benchtop Hettich Rotina 380R for general purpose applications, minimum tube size 0.2ml, maximum tube size 290ml, maximum g force 24400, min temperature -20°C, max temperature 40°C, run counter, no rotor included

read more

List Price: Offer Price: Quantity:
£8,901.00 £6,199.00
List Price: £8,901.00
Offer Price: £6,199.00
Quantity:
Manufactured in european2 Years WarrantyDelivery Time

Centrifuge benchtop Hettich Rotina 380R for general purpose applications, minimum tube size 0.2ml, maximum tube size 290ml, maximum g force 24400, min temperature -20°C, max temperature 40°C, run counter, no rotor included


Power Draw:1.30kW
Ecological Features:Opening the lid of refrigerated units will turn off the refrigeration so it will not use lots of energy when left on.
Min tube volume:0.2ml
Max tube volume:290ml
Max g force:24400
Min temperature:-20°C
Max temperature:40°C
For general purpose use:Yes
For cell culture - low capacity use:No
For cell culture - medium capacity use:No
For cell culture - high capacity use:No
For cytology - CSF use:No
For cytology - urine use:No
For microplates:No
For serum/plasma use:No
Run counter:Yes
Rotor included:No

Minimum specification that this product meets

  • A motorised lid interlock system that prevents user access to the rotor whilst it is in motion and prevents the rotor turning whilst the lid is open
  • The rotor bowl, lid and external casing must be sufficiently robust to be able to contain the rotor and all samples therein in the event a rotor or drive spindle failure
  • A range of interchangeable rotor options, including swing-out or fixed-angle rotors, with a variety of bucket/tube capacities.
  • Rotor imbalance detection and cut out
  • A variable speed controller with the potential of centrifugation up to 20,000rpm although it is realised that some rotors will not be capable of this speed. The accuracy of the rotor speed against the display should be no more than
  • 100rpm or
  • 1%, whichever is the least.
  • A facility to control the acceleration/deceleration rates of the rotor.
  • An adjustable timer to control the duration of centrifugation with a minimum range of 1 minute to 99 hours
  • A cooling system that does not utilise CFC refrigerants
  • Adjustable rotor and rotor bowl temperature with a minimum range of 0oC to +40oC in 1oC increments and an variation not exceeding
  • 2oC
  • Straightforward, self-evident controls requiring the minimum of training
  • Legible digital display(s) capable of showing set time, temperature and speed and actual time, temperature and speed
  • Electrical operation utilising a plug in or hard-wired isolated 30 Amp, single phase supply at 230V 50Hz or from a 230 volt, AC (50hz), power supply with a three-pin plug top to BS1363