Item shown is representative of range - Catalogue No.:AMB29

Item shown is representative of range - Catalogue No.:AMB29

Rodwell - Ambassador 158DS

Stock Code: AMB29

Wolf special offer - Autoclave Rodwell Ambassador 158DS user programmable, suitable for liquids, no vacuum air removal, no autofill, 158l capacity, usable depth 960mm, usable diameter 460mm, no data output OFFER ENDS 24-06-2021

read more

List Price: Offer Price: Quantity:
£16,617.00 £14,129.18
List Price: £16,617.00
Offer Price: £14,129.18
Quantity:
Manufactured in uk1 Year WarrantyDelivery Time

Wolf special offer - Autoclave Rodwell Ambassador 158DS user programmable, suitable for liquids, no vacuum air removal, no autofill, 158l capacity, usable depth 960mm, usable diameter 460mm, no data output OFFER ENDS 24-06-2021


Power Draw:3.600kW
Type:Floor standing
Capacity:158l
Suitable for liquids?:Yes
User programmable:Yes
Loading:Front
Autofill:No
Vacuum air removal:Optional
Number of chambers:1
Heat source required:No
Chamber shape:Round
Usable width:0mm
Usable height:0mm
Usable depth:960mm
Usable diameter:460mm
Heating method:From steam supply
Max run temperature:138°C
Automatic door opening:No
Sliding door:No
Temperature and pressure gauges included:Yes
Data output:Optional - see accessories
Disinfection type:Steam
Assisted air cooling:No
Suitable for laboratory use:Yes
Suitable for dental use:No
Suitable for medical use:No
Suitable for beauty/tattoo use:No

Minimum specification that this product meets

  • The chamber to be constructed from high-grade stainless steel conforming to PD5500:2009 as amended
  • The general design, construction, operation and performance should conform to BS2646-1:1993 as amended.
  • Capable of generating its own steam electrically.
  • Chamber access via a door fitted with a safety mechanism that allows any residual pressure in the chamber to vent safely before the door can be opened fully.
  • Control panel, incorporating a chamber pressure gauge, allowing the user to select appropriate sterilisation cycles and monitor the progress of the sterilisation cycle via a digital/graphical display.
  • A minimum of 4 user selectable cycle programmes held in the autoclaves memory
  • The facility to lock cycle settings to prevent inadvertent changes to the settings by infrequent users. A key or PIN accessed supervisor mode would be necessary to set the parameters for the cycles.
  • User adjustable sterilisation temperatures in the range 100oC to 137oC. User adjustable sterilisation times, activated by the achievement of the preset sterilisation temperature, in the range of 1
  • 60 minutes.
  • User adjustable free-steaming time in the range of 0
  • 30 minutes
  • A thermal lock, with key override, on the chamber door that prevents opening before the chamber and goods cool to 80oC