Item shown is representative of range - Catalogue No.:SQ-4215

Item shown is representative of range - Catalogue No.:SQ-4215

SciQuip - Incu-Shake FL18-750R

Stock Code: SQ-4215

Shaking incubator SciQuip Incu-Shake FL18-750R floor standing, refrigerated, max vessel capacity 5000ml x 1, front loading, 26mm orbit, not stackable, includes 496 x 350mm platform. Please note that the upper platform must be removed for use with 200...

read more

List Price: Offer Price: Quantity:
£10,141.00 £5,512.00
List Price: £10,141.00
Offer Price: £5,512.00
Quantity:
Manufactured in worldwide2 Years WarrantyDelivery Time

Shaking incubator SciQuip Incu-Shake FL18-750R floor standing, refrigerated, max vessel capacity 5000ml x 1, front loading, 26mm orbit, not stackable, includes 496 x 350mm platform. Please note that the upper platform must be removed for use with 2000ml, 3000ml and 5000ml flasks, max shaking capacity 13500ml


Power Draw:0.48kW
Benchtop or floor standing:Floor standing
Refrigerated:Yes
Max vessel capacity:5000ml
Platform access:Front loading
Max orbit:26mm
Programmable:No
Stackable?:No
Block or platform included?:Yes
Type of shaking:Orbital
Max shaking capacity:13500ml
External width:720mm
External depth:630mm
Minimum temperature:4°C
Maximum temperature:60°C`
Maximum RPM:300RPM
With antimicrobial protection:No

Minimum specification that this product meets

  • Operate at a Sub-ambient temperature, from 15oC below ambient to 60oC
  • A maximum variance across all areas of the incubator chamber of ?0.5oC at any set-point of 20oC (type 1) and 37oC (type 2)
  • An accuracy of ?0.1oC of the displayed temperature
  • An adjustable orbital platform speed control between 25rpm and 400rpm with a deviation of speed of ?1% of the set point
  • An easily accessible and simple to use controller with digital set points and temperature and orbital speed displays.
  • An audible and visible, adjustable high and low temperature alarm
  • A range of interchangeable orbital platform with clamping devices to accommodate range of sizes of vessels
  • Door opening mechanisms that require the minimum of effort and incorporates a safety interlock that stops the orbiting platform when open
  • Overload sensors on the orbital platform to make user aware of overload condition and restrict orbiting speed to a safe level
  • All motor and platform bearings are to be sealed to prevent ingress of liquid in the event of a spillage
  • CFC free refrigerants on sub-ambient models
  • Electrical operation from a 230 volt, AC (50hz), power supply with a three-pin plug top to BS1363