Item shown is representative of range - Catalogue No.:HR900-IIA2

Item shown is representative of range - Catalogue No.:HR900-IIA2

Haier Biomedical UK - HR900-IIA2

Stock Code: HR900-IIA2

Class II microbiological safety cabinet Haier Biomedical HR900-IIA2 working width 920mm, manual sliding sash, one fan, 0.4kW max power draw, 60dB noise level, no VHP port, for recirculating applications, with angled HEPA filter and 1 year waranty, in...

read more

List Price: Offer Price: Quantity:
£11,041.00 £7,051.00
List Price: £11,041.00
Offer Price: £7,051.00
Quantity:
Manufactured in worldwide1 Year WarrantyDelivery Time

Class II microbiological safety cabinet Haier Biomedical HR900-IIA2 working width 920mm, manual sliding sash, one fan, 0.4kW max power draw, 60dB noise level, no VHP port, for recirculating applications, with angled HEPA filter and 1 year waranty, includes stand


Power Draw:0.40kW
Ecological Features:DC and EC fans and longer lasting LED lights reduce energy consumption.
Working width:920mm
Front closure:Manual Sliding Sash
Number of fans:1
Max power draw:0.4kW
Noise level:60dB
VHP ports?:Optional
Ducted or recirculating:Recirculating
Angled HEPA filter?:Yes
Side Walls:Glass
Number of access ports:2
Warranty:1Years
Working height:650mm
Working depth:620mm

Minimum specification that this product meets

  • Compliance with the requirements of BS5726:2005/EN 12469:2000 and other current, relevant standards
  • Stainless steel (minimum grade AISI 316) or epoxy coated work surfaces
  • MSC to have conspicuous visual indication that the unit is operating safely and an audible & visual alarm indicating any condition that may render the MSC unsafe to use.
  • Air flow rates of at least 0.4 metres/second inflow (average across the face) and a down flow over the working area of at least 0.25 metres/second but no greater than 0.5 metres/second.
  • A variety of widths
  • An inlet to allow for oil aerosol penetration tests of the HEPA filters and seals
  • An additional HEPA exhaust filter (non-ducted models only)
  • Internal lighting in the workspace of at least 1,000 Lux
  • A site visit by a trained operative to advise on the suitability of the proposed location of the MSC.
  • Electrical operation from a 230 volt, AC (50hz), power supply with a three-pin plug top to BS1363