 | Order. no. 10855 (3 x 400 V /50 Hz) air cooled
Order. no. 10856 (3 x 220 V / 60 Hz) air cooled
Order. no. 10857 (3 x 400 V / 50 Hz) water cooled
|
 | Compact refrigerated floor standing centrifuge for universal use in
blood banks, clinical laboratories and research
|
 | Capacity up to 6 l in angle rotor; 12 x 1 l, 6 x 1.5 l or 12 blood
bags in swing-out rotor
|
 | Solid construction, suitable for continuous use
|
 | Maintenance-free drive, high imbalance tolerance up to approx. 80 g
|
 | Speed range 100 - 10.500 rpm
|
 | Automatic rotor identification prevents spinning rotors overspeed
|
 | Efficient refrigeration; temperature adjustable between -20 °C and
+40 °C, standstill cooling, with refrigerant R404a
|
 | Stainless steel chamber
|
 | The lid is hinged on the left hand side
|
 | Easy open the centrifuge lid via pneumatic springs
|
 | Two motorized lid locks provide safe, assured lid-locking
|
 | Power failure, emergency lid opening
|
 | Ergonomic loading height of just 88 cm
|
 | The user friendly control unit is integrated in the front panel
|
 | Centrifuge fits through a door opening of only 82 cm
|
 | The centrifuge stands on movable castors an adjustable feet to even
out the floor
|
 | Easily serviceable
|
 | All international standards are met, e.g. IEC 1010, CE, WEEE
|
Options
 | Data logging via RS 232
|
 | Water-cooled compressors
|
 | Sealed housing and water-cooled compressors, sterile room version
|
 | Inert gas injection with analysis of the oxygen content in the
exhaust emission |
Spincontrol - Professional
 | Sophisticated version of the unique SIGMA single-knob control with
backlit graphical LCD-display
|
 | Selectable languages: German, English
|
 | All important run parameters at a glance
|
 | Permanent display of set and actual values: time, speed,
gravitational field and temperature
|
 | 60 programs can be named and stored in a memory with all run
parameters
|
 | Rapid cooling program
|
 | 10 linear, 10 quadratic and 10 user defined acceleration an
deceleration curves including brakeless deceleration
|
 | Setting of values in steps of 100 or 1 rpm resp. 10 or 1 x g
|
 | Time 0 - 99 h 59 min, continuous operation, short run
|
 | Electronic imbalance monitoring, acoustic and visual alarm with fast
deceleration
|
 | End of run alarm, selectable, can be switched of
|
 | Plain text display of information an error messages
|
 | Run time and cycles count for the rotors used
|
 | Selectable additional functions:
Time integral, standstill cooling, automatic unlocking of lid after end
of run, timing after set speed has been reached, password protection of
programs in memory, information in case of excessive chamber temperature
with adjustable tolerance, speed adjustment for substances with density
exceeding 1,2 g/ccm, start delay, automatic speed adjustment for
different radii to have same gravitational field, temperature display in
°F or °C, display of acceleration and deceleration curves, program
rotation
Option:
Interface for optical end of run indication
|
|