- Automatic rotor identification system
- The HARRIER range incorporates an automatic rotor
identification system. This important feature ensures that
the centrifuge recognises the rotor in use and prevents it
from being used beyond its safe maximum speed. As the
centrifuge identifies the rotor in use, it can also work out
the ‘g’ force at any set speed. This means the centrifuge
can convert rpm to ‘g’ and vice versa once in use.
- Microprocessor controlled induction drive
- The HARRIER range utilises an induction motor for
maintenance free long life operation. This motor has no
carbon brushes so not only is there no need for routine
maintenance but there is no build-up of carbon dust inside
the centrifuge. This makes it suitable for use in clean
environments. Microprocessor control combined with the
frequency driven motor ensures accurate operation for run to
run reproducibility.
- Refrigerated version available
- The HARRIER 18/80 REFRIGERATED combines all the
capabilities of the proven HARRIER 15/80 in a refrigerated
format. The selectable temperature range is from −90C to
+40C and uses CFC–free R134a refrigerant gas. This means
that in refrigerant terms, the HARRIER 18/80 REFRIGERATED
complies with all the regulations pertaining to the Montreal
Protocol. A pre-cool facility allows the bowl to be cooled
to the required temperature prior to use.
- High speed microtube rotor
- The use of the induction motor provides another benefit
in that higher rotational speeds can be achieved. This
allows the HARRIER range to have a maximum speed in excess
of 15,000 rpm with the 24 place microtube rotor. The
addition of refrigeration in the HARRIER 18/80 REFRIGERATED
unit allows the microtube rotor an increase in speed to
18,000 rpm giving a maximum ‘g’ force of 28,980 ‘g. See
rotor table for details.
- Wide range of accessories available
- For routine bench top work, the 4 x 200ml swing out
rotor and 200ml sealed buckets are available on the HARRIER
15/80. A recent addition on the HARRIER 18/80R is the 4 x
200ml wind-shielded rotor. This uses the same 200ml buckets
as used on the HARRIER 15/80 but with an increased speed
capability of 6,000 rpm giving a maximum g force of 6,466 g.
These, combined with a wide range of adaptors, can
accommodate most manufacturers’ tubes. As well as the swing
out rotor, other angle rotors and specialist rotors are
available. Please see tables for details.
|
|