FOREWORD
is a compact, economical new concept in vertical
laminar airflow workstations. MINILAB uses very little bench
space, yet provides a decontaminated bench top work space in
which to carry out manipulations which must be kept clean.
MINILAB is easy to move and can also be placed in different
laboratories.
APPLICATIONS
MINILAB can be used in a wide range of disciplines
including:
- Microbiology
- Sterile manipulations
- Cell culture
SPECIFICATIONS
Construction:
corrosion resistant epoxy powder painted steel.
Work surface: glazed
AISI-304 perforated stainless steel -Removable for
autoclaving
Transparent side walls and front
windows: plexiglass.
Back side: opaque
plexiglass.
Absolute filter: HEPA
filter, tested with D. 0. P. aerosol. Their efficiency,
higher then 99.999% at 0.3 micron particle size - 0.001%
penetration - guarantees performance exceeding requirements
of U. S. Federal Standard 209 e, BS 5295, VDI-2083 and AS
1386 -1976.
Prefilter:
syntethic-fibers pre-filter (washable in water or with
compressed air), average arrestance 80/90% ASHRAE (CEN-G3)
dust holding capacity 400 g/m2.
Motor blower: direct coupled motor, electronic
speed controlled to maintain a constant air flow of 0.45
m/sec, and compensate for a partially clogged filter up to a
maximum plenum pressure of 35 mm of water.
Power Supply: 220 V, 50
Hz, single phase. Other voltages available on request. |
| |
|
Code
|
Model
|
Useful
Dimensions mm
|
Overall
Dimensions mm
|
Power
|
Volts
|
Weight
|
Flow
Rate
|
| |
|
W
|
H
|
D
|
W
|
H
|
H
|
Kw
|
Hz
|
Kg
|
m3/h
|
| F72 799900 |
MINILAB |
580
|
375
|
317
|
642
|
830
|
340
|
0.11
|
220/50 |
50
|
290
|
|
|