FOREWORD
BH-EN 2000 Class II Microbiological Safety
cabinets belong to the latest generation of laminar airflow
systems, combining rigid safety requirements with high
quality construction. BH-EN 2000 has been designed as a
Class II Microbiological Safety cabinets according to
EN-12469. Therefore it provides vertical laminar airflow
with part of the air recirculated via the main HEPA filter
whilst the remaining part is discharged through an exhaust
HEPA filter. This cabinet is suitable for handling CDC 1-3,
DPAG C-B1-B2 etiologic agents, providing operator, product
and environmental protection
APPLICATIONS
BH-EN 2000 cabinets can be used in a wide range of
disciplines, as defined in appropriate standards. Especially
suitable for applications such as:
- Microbiology
- Virology
- Haematology
- Cell culture
- Recombinant DNA
- Handling of hazardous agents to human beings or animals as
defined in the appropriate international standards.
BENEFITS
- Sophisticated microprocessor based monitoring system
Sophisticated microprocessor based system of audible and
visual alarms alerts the operator if the cabinet does not
operate in complete safety.
- Self-adjusting system to ensure a constant laminar airflow
speed even in the presence of a progressive clogging of the
filters.
- The negative pressure plenum "bio-dynamic sealing" ensures
that all contaminated particles are kept inside the system
and are automatically drawn to the pressure chamber to be
captured and by the main and exhaust filters.
- The fan system assures that no part of the cabinet is
under positive contaminated pressure to the laboratory, thus
preserving environment and personnel from bio-contamination
risks.
- Access from the front panel for ease of maintenance
service.
- Available in one motor (BH-EN
2000 S) as well as two motors (BH-EN 2000 D) versions for
flexible performance selection in line with the cabinet's
use.
SPECIFICATIONS
Construction:
Epoxy powder painted steel for
improved strength and resistance to corrosion and
atmospheric agents.
Work Surface: 3,
4, 5 or 6 pieces (depending on the cabinet width) AISI-316L
stainless steel, completely perforated to ensure flow
laminarity and air recycling at the work surface level.
Frontal slots in the access area provide air barrier
protection preventing any exchange of air between the
inside of the cabinet and the environment.
On request the following work surfaces are available with no
extra charges:
- AISI-316L stainless steel perforated work surface in 1
piece
- AISI-316L stainless steel solid work surface in 1 piece or
3, 4, 5, 6 pieces (depending on the cabinet width).
Work Chamber: in
AISI-304 stainless steel sheet.
Absolute Filter:
HEPA filters, tested with D.O.P. aerosol. Their efficiency,
higher than 99,995%MPPS - 0.005% penetration - ensures a
performance exceeding the requirements ofEN 1822, U.S.
Federal Standard 209 e, British Standard 5295, VDI-2083 and
AS 1386-1976.
The filtration system is designed to provide a barrier of
air with minimum average speed over 0.4m/sec. according to
EN-12469.
Motor-blower/s:
One centrifugal fan (BH-EN 2000 S) and two centrifugal fans
(BH-EN 2000 D), direct driven motors, with protection factor
IP-55, controlled by two automatic speed regulators and
microprocessor system showing laminar flow and protection
barrier work conditions.
Microprocessor based monitoring
system: new generation system for
controlling the cabinet in a full digital way performs high
level control and data processing. The handy and practical
keyboard and the rearlit LCD display keep the user
constantly informed of the cabinet conditions and operation,
in particular:
- display of laminar airflow velocity and frontal air
barrier velocity
- display of inside and outside temperature
- display of residual lifetime of HEPA filters, U.V. Lamp
and charcoal filter (if installed)
- display of total time of cabinet operation
- display of power rating factor of the main and exhaust (if
installed) motor-blowers.
Alarms for:
- out of range laminar airflow velocity and front air
barrier velocity
- front window opened.
- HEPA filters clogging
- UV lamp worn-out (if installed)
- charcoal filter (if installed) saturated
- blackout
- signal of other possible malfunctions and relevant
remedies.
Power Supply:
230V, 50Hz, single phase.
Lighting:
fluorescent tubes in built-in housing, placed outside the
contaminated area.
Front window:
Stratified safety glass, to give ease of access for large
items. It is provided with gas spring to keep it open during
maintenance or sanitization operations.
Additional Features:
- two vacuum or gas connections in the working area
are provided, one with manual tap, the other with solenoid
valve (on the right-hand side).
- A stainless steel spillage tray beneath the work surface
collects spilt liquids.
- A stainless steel front closure panel ensures the sealing
of the cabinet for sterilization/fumigation .
- An inlet allows the DOP test of the absolute HEPA filters.
- Electrical socket with protection IP-44 (on the right-hand
side)
Maintenance: access to filters, motor-blower/s and control
panel by lifting the front panel.
As Option:
- UV lamp for sterilization of the work chamber
installed onthe front closure, controlled by two timers, one
with 0¸3 hours scale (1 minute steps), the other 3 fixed
hours. |
|
Code
|
Model
|
Useful
Dimensions mm
|
Overall
Dimensions mm
|
Power
|
Weight
|
| |
|
W
|
H
|
D
|
W
|
H
|
D
|
Kw
|
Kg
|
|
F74 600100
|
BH-EN 2003-S
|
885
|
660
|
580
|
1090
|
1470
|
785
|
0.7
|
181
|
|
F74 600120
|
BH-EN 2004-S
|
1190
|
660
|
580
|
1395
|
1470
|
785
|
1
|
213
|
|
F74 600130
|
BH-EN 2005-S
|
1495
|
660
|
580
|
1700
|
1470
|
785
|
1.2
|
260
|
|
F74 600140
|
BH-EN 2006-S
|
1800
|
660
|
580
|
2005
|
1470
|
785
|
1.4
|
298
|
|
F74 600210
|
BH-EN 2003-D
|
885
|
660
|
580
|
1090
|
1470
|
785
|
0.85
|
186
|
|
F74 600220
|
BH-EN 2004-D
|
1190
|
660
|
580
|
1395
|
1470
|
785
|
1.1
|
218
|
|
F74 600230
|
BH-EN 2005-D
|
1495
|
660
|
580
|
1700
|
1470
|
785
|
1.3
|
265
|
|
F74 600240
|
BH-EN 2006-D
|
1800
|
660
|
580
|
2005
|
1470
|
785
|
1.5
|
304
|
|
FrontAperture
|
Temperature rise
|
Exhaust duct
|
Noise level
|
Lighting
|
Vibrations
|
|
200mm
|
< 4°C
|
Ø200 mm
|
<59 dBA
|
>1000 lux
|
<0.005 mm RMS
|
 |
The model BH-EN 2004 D
is certified from TUV NORD according to EN12469
|
|
Certification Number:
08/205/GS/YZW140899 |
| |
|
 |
The models BH-EN 2003
S, BH-EN 2003 D, BH-EN 2004 S,BH-EN 2004 D are
certified from LNE according to EN12469 |
|
Certification Number: NF
095 01/01/02/01-06 |
| |
|
 |
|
|